Descargar Imprimir esta página

Installing The Timer; Polarity Test; Mounting Of The Tubehead; Closing The Unit - GENDEX Oralix AC Manual De Servicio

Ocultar thumbs Ver también para Oralix AC:

Publicidad

Idiomas disponibles

Idiomas disponibles

INSTALLATION

3.5. INSTALLING THE TIMER

Remove the cover of the Timer and fit it on
1
the two screws stud (b) provided on the
mounting plate of the column (see figure).
2
Fix it with the nuts and rings supplied.
Connect on the terminal block of the Timer
3
the wires from the Folding Arm cable and the
wires of the cable from the column to the
Timer's terminal block, according to BZ-5.
Ensure that the protective earth lead (y/g) from
Folding Arm cable is securely connected at
block terminal pin X3:1; the protective earth
lead (y/g) from mains power of the column is
securely connected at block terminal pin X3:3;
the protective earth lead (y/g) from the stud
of the column is securely connected at block
terminal pin X3:3.
Verify the jumper and dip-switches settings
4
as described in Section C,"Jumpers and Dip-
switches setup" of this manual.
Do not remount the Timer cover until the po-
larity check has been performed.

3.6. POLARITY TEST

Please refer to paragraph "2.5 Polarity test" on this manual.

3.7. MOUNTING OF THE TUBEHEAD

Please refer to paragraph "2.6 Mounting of the tubehead" point 2 and following on this manual.
3.8

CLOSING THE UNIT

Before closing the unit perform the Setting to work actions, as described in the section C of this manual.
B-20
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER
(b)
ORALIX AC SYSTEM
(b)
Gendex Dental Systems

Publicidad

loading

Productos relacionados para GENDEX Oralix AC