Important Information For Mounting And Maintenance; Electrostatic Charging (Esd) - Vega VEGAPULS C 21 Manual De Usuario

Ocultar thumbs Ver también para VEGAPULS C 21:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

7
Important information for mounting and maintenance
General instructions
The following requirements must be fulfilled for mounting, electrical installation, setup and mainte-
nance of the instrument:
The staff must be qualified according the respective tasks
The staff must be trained in explosion protection
The staff must be familiar with the respectively valid regulations, e.g. planning and installation
acc. to IEC 60079-14
Make sure when working on the instrument (mounting, installation, maintenance) that there is no
explosive atmosphere present, the supply circuits should be voltage-free, if possible.
The instrument has to be mounted according to the manufacturer specifications, the Certificate
of Conformity and the valid regulations and standards
Modifications on the instrument can influence the explosion protection and hence the safety,
therefore repairs are not permitted to be conducted by the end user
Modifications must only be carried out by employees authorized by VEGA company
Use only approved spare parts
Mounting
Keep in mind for instrument mounting
Mechanical damage on the instrument must be avoided
Mechanical friction must be avoided
If the device is used as a separating wall device, the operator must observe the applicable
installation regulations.
Maintenance
To ensure the functionality of the device, periodic visual inspection is recommended for:
Secure mounting
No mechanical damages or corrosion
Worn or otherwise damaged cables
No loose connections of the line connections, equipotential bonding connections
Correct and clearly marked cable connections
8
Electrostatic charging (ESD)
Take note in case of danger of electrostatic charges:
Avoid friction on the surfaces
Do not dry clean the surfaces
The instruments must be mounted/installed in such a way that the following can be ruled out:
electrostatic charges during operation, maintenance and cleaning.
process-related electrostatic charges, e.g. by measuring media flowing past
The warning on the type label indicates danger:
WARNING - POTENTIAL ELECTROSTATIC CHARGING HAZARD – SEE INSTRUCTIONS
AVERTISSEMENT – DANGER POTENTIEL DE CHARGES ELECTROSTATIQUES – VOIR
INSTRUCTION
For media with a conductivity smaller than 10
– The level measuring instrument must not be used in highly charge generating processes, e.g.
mechanical friction and separation processes, spraying of electrons, etc.
VEGAPULS C 21, C 22, C 23
S/m applies:
-8
IECEx
5

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Vegapuls c 22Vegapuls c 23

Tabla de contenido