EN
User/Patient Safety
WARNINGS:
• Only trained and experienced healthcare
professionals should use this equipment.
Before using any system component
or any component compatible with
this system, read and understand the
instructions. Pay particular attention to
WARNING information. Become familiar
with the system components prior to use.
• Upon initial receipt and before each
use, inspect non-sterile batteries for
damage. DO NOT charge or use if there
is evidence of damage. Electrolytes may
leak from damaged batteries and cause
serious chemical burns to the surgical
site, skin and/or eyes.
• DO NOT modify any system component or
accessory.
4
Print Date: Jun 14, 2016 11:18:38 AM
0000003946, Rev. L Effective Date: Jun 14, 2016 10:36:00 AM
• DO NOT use this equipment in areas in
which flammable anesthetics or flammable
agents are mixed with air, oxygen or
nitrous oxide.
• See the instructions supplied with the non-
sterile battery and the System 5 or System
6 Charger used to recharge the batteries.
Instructions contain important safety
information and charging instructions.
• ALWAYS remove the non-sterile battery
from the aseptic housing before charging
the battery. Failure to comply may result in
product damage.
• Sterilize the aseptic housings and transfer
shields before first and every use. See the
care instructions manual supplied with the
handpiece.
• DO NOT sterilize the non-sterile batteries.
After charging, a non-sterile battery is
placed in a sterilized aseptic housing, then
loaded into a handpiece.
4126-120-700 Rev-L
www.stryker.com