Set-up
The unit must be placed on a stable, flat, level surface. Allow sufficient space
around the unit for safe operation.
Take care that the turning rotor cannot collide with anything else on the
laboratory table. Operating personnel have to be aware of the risk of
accidents and observe safety rules.
2
Electric hook-up
Connect mains plug.
3
Bottle installation
3.1 2000 ml bottles
The standard version of the unit is designed for 2000 ml
bottles. Set the rotor in the proper position in order to install
the bottles.
To do this, switch (A) on and set speed control knob (B) to 1
rpm. Turn switch (A) off when levers (C) of one side are in
the upright position.
Follow the instructions below and then repeat the entire
process for the other side of the rotor.
SET-UP
Release lever (C) by turning it anticlockwise. Pull up on
button (D) in order to raise the bottle holder upwards.
Insert the bottle and take care that the bottle holder
centers exactly above the screw cap of the bottle.
Push button (D) down in the direction of the bottle.
Tighten lever (C) clockwise.
Repeat the procedure according to the number of bottles.
Check the stability of the bottles and re-tighten them, as
described above, if necessary.