Electromagnetic Compatibility Guide - Lepu Medical LOX100A Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Anti-shock protection category
Anti-shock protection measure
Waterproof protection measure
Net Weight
Dimensions
Operating mode
Spo2
Data averaging
Pulse rate
Spo2
Update time
Pulse rate
13.Electromagnetic Compatibility Guide
Note:
·This device should not be used close to or stacked with other devices.
If it must be used close to or stacked with other devices, care should be
taken to verify that it functions properly under its intended use.
·Except for the cables of this product sold by the manufacturer as
spare parts for internal components, use of the accessories and cables
other than those specified may result in increased emission or reduced
immunity of this product.
·Since portable and mobile RF communication equipment may affect
the performance of this product, please avoid strong electromagnetic
interference during use, such as mobile phones, microwave ovens and so
on.
·The user should install and use the device according to the
electromagnetic compatibility information provided in the random file.
Internal power supply
Type BF applied part
IP22
approx. 60 g (including batteries)
69 mm (L) x 35 mm (W) x 29 mm (H)
Non-continuous operation
Average of successive five detected
pulses, with exponential smoothing
followed
Average within 8 seconds
Update per second, the update
period is less than 20 seconds.
Update per second, the update
period is less than 12 seconds.
13
(1)
Guide and manufacturer's statement - Electromagnetic emissions
This product is intended for use in the following electromagnetic
environment.The purchaser or user of theproduct should ensure that it
is used in this electromagnetic environment
Emission test
Compliance
RF emissions
Group 1
RF emissions
Class B
Harmonic
Not applicable
emissions
Voltage
fluctuation
Not applicable
/ flickering
emissions
(2)
Guide and manufacturer's statement - Electromagnetic immunity
This device is intended for use in the following electromagnetic
environment. The purchaser or user of this device should ensure that it
is used in this electromagnetic environment
IEC60601 test
Immunity test
level
± 8 kV contact
Electrostatic
discharge
discharge
± 15 kV air
discharge
Electromagnetic environment - Guide
This product uses RF energy only for
its internal functions. As a result, its
RF emissions are low and there is
very little chance of interference with
nearby electronic equipment.
This product is intended for use in all
facilities including domestic facilities
and the facilities connected directly
to public low-voltage power supply
network for residential homes.
Compliance
Electromagnetic
level
environment - Guide
The floor should be
wood, concrete or
± 8 kV contact
ceramic tile; if the
discharge
floor is covered with
± 15 kV air
synthetic materials,
discharge
the relative humidity
should be at least
30%.
14

Publicidad

Tabla de contenido
loading

Productos relacionados para Lepu Medical LOX100A

Este manual también es adecuado para:

Lox100bLox100cLox100d

Tabla de contenido