POTENTIAL FOR ELECTROMAGNETIC
INTERFERENCE
To avoid inaccurate results caused by electromagnetic interference
between electrical and electronic equipment, do not use the device near
a cell phone or microwave oven. For most wireless
communication devices, it is recommended to
maintain a distance of 10.8 feet (3.3m) in order to
avoid electromagnetic interference.
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions:
(1) this device may not cause harmful interference,
and (2) this device must accept any interference
received, including interference that may cause
undesired operation.
NOTE: The manufacturer is not responsible for any radio or TV
interference caused by unauthorized modifications to this equipment.
Such modifications could void the user authority to operate the equipment.
NOTE: This equipment has been tested and found to comply with the
limits for a Class B digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses
and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will
not occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct
the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from
that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
31
ELECTROMAGNETIC COMPATIBILITY (EMC)
• Guidance and manufacturer's declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should only be
used in such environments:
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
RF energy is used only to maintain device's operation.
Therefore, its RF emissions are so low that it's not
RF emissions CISPR 11
Class B
likely to cause any interference in nearby electronic
equipment.
Harmonic emissions IEC
Not
The device is suitable for use in all establishments,
61000-3-2
Applicable
including domestic establishments, and those directly
connected to the public low-voltage power supply
Voltage fluctuations/
Not
network that supplies buildings used for domestic
flicker emissions IEC
Applicable
purposes.
61000-3-3
• Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be
used in such environments:
Immunity test
IEC 60601 test level
Compliance level
Electrostatic
± 6 kV contact
± 6 kV contact
discharge (ESD)
± 8 kV air
± 8 kV air
IEC 61000-4-2
Power frequency
3 A/m
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
• Recommended separation distances between portable and mobile RF communication equipment
and the device.
The device is intended for use in an electromagnetic environment where radiated RF disturbances are
under control. User can help prevent electromagnetic interference by keeping the device at a minimum
distance from portable and mobile RF communications equipment (transmitters). Below table details
the maximum output power of transmitter:
Rated maximum
Separation distance according to frequency of transmitter m
output power of
150 kHz to 80 MHz
80 MHz to 800 MHz
transmitter W
Not Applicable
d = 1.2
0.01
NA
0.1
NA
1
NA
10
NA
100
NA
32
Electromagnetic environment–
guidance
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.
800 MHz to 2.5 GHz
d = 2.3
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23