Hydrazine P / M205
Hydrazine P
0.05 - 0.5 mg/l N
Dimethylaminobenzaldehyde
Material
EN
Required material (partly optional):
Reagents
Hydrazine Test Powder
The following accessories are required.
Accessories
Measuring spoon, 1 g
Preparation
1. If the water sample is turbid, it must be filtered before performing the zeroing.
2. The sample's temperature should not exceed 21 °C.
Notes
1. When using the hydrazine measuring spoon, 1 g is a level measuring spoon.
2. For removal of the reagents resulting in turbidity, ensure to use a quality membrane
filter for medium deposits.
3. To check the reagent for prolonged storage and possible ageing, follow the test as
described for tap water. Should the result of the value of the detection limit of 0.05
mg/l be exceeded, the reagent may only be used with restrictions (larger measured
value deviations).
26
H
2
4
Packaging Unit
Powder / 30 g
Packaging Unit
1 pc.
EN   Method Reference Book   1.0
M205
Hydr
Part Number
462910
Part Number
384930