Implementation of the provision Oxygen, dissolved with Vacu
Vials
K-7553
®
Select the method on the device
Place Zero ampoule in
the sample chamber.
Run test water through the
sampling vessel for several
minutes from bottom to top
to remove air bubbles.
Close the opening with one
finger, to avoid contact with
the air.
EN   Method Reference Book   1.0
Zero
Press the ZERO button.
PRESS
Place a Vacu-vial
ampoule
®
in the sampling vessel.
Break off the ampoule tip
by applying light pressure
against the vessel wall.
Wait for the ampoule to fill
completely.
Invert the ampoule several
times.
Oxygen dissolved C / M292
Remove zero ampoule
from the sample chamber.
Then quickly remove the
ampoule from the sampling
vessel with the tip down.
Dry the outside of the
ampoule.
EN
39