Descargar Imprimir esta página

Fig. 5-4 Scale And Pump Connection - Hologic Aquilex Instrucciones De Uso

Ocultar thumbs Ver también para Aquilex:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 190
System Operation
en
Fig. 5–4  Scale and pump connection
(1)
Bag deflector
(2) Power cord/pump connection
(3) Covered power output
(4) Plug connection, bag scale with
container scale
(5) Bag scale connector, connection
to pump
(6) Handwheel
(7) Screw
24 / 378
5.
Guide the power cord through the holes provided for this purpose and connect
to pump (2) (Fig. Scale and pump connection [} 24]) and plug into grounded,
shockproof safety wall socket.
6.
Attach the scale to the pump by connecting and screwing the connector of the
bag scale together with the connector of the container scale (4) (Fig. Scale and
pump connection [} 24]) and fix the connected cables below the lower pump
tray by means of the provided wire clips. Connect and screw the second con-
nector of the bag scale to the port on the back of the pump (5) (Fig. Scale and
pump connection [} 24]).
WARNING!
Scale error
Ensure that nothing weighs down the scale during system start-up. Doing so may
result in an inaccurate deficit value.
WARNING!
Fluid deficit
The fluid left in the patient must be monitored. The deficit is the total amount of
fluid left in the patient or unaccounted for otherwise. Take notice of the measure-
ment tolerance of the system (see Chapter  Technical Data [}  49]). Estimating the
fluid volume remaining in the patient is the physician's responsibility.
WARNING!
Serum sodium concentration
It is also necessary to monitor the concentration of sodium in the blood of the pa-
tient to prevent electrolyte disturbances. Monitoring of the concentration of so-
dium in the blood must be performed by the physician and is not performed or sup-
ported by the system.
(1)
(2)
(3)
(5)
(6)
(7)
(4)

Publicidad

loading