Calibrating The Gauge; Locking & Unlocking The Calibration - Elcometer 456 Guia Del Usuario

Ocultar thumbs Ver también para 456:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 35
6 TAKING A READING (continued)
DO NOT
û
Drag the probe over the coated surface.
Bang the probe down hard onto the surface.
Allow the probe to hover over the surface as this could result in
a false reading.
4
The display will dim if inactive for more than 15 seconds and will go 'black' if
inactive for the period defined in Menu/Setup/Screen Settings/Screen Timeout.
Either press any key, or tap the gauge to awaken it.
4
The gauge will switch off automatically after 5 minutes of inactivity.
4
- - - indicates reading outside range of probe.

7 CALIBRATING THE GAUGE

1
Press the Cal softkey.
2
For alternative calibration methods,
select Cal/Cal Method.
3
Select Calibrate and follow the on-
screen instructions.
4
When prompted place the probe on
the centre of the foil(s).
4
Foils can be stacked.
4
Not all calibration methods are available for all gauge types see 9 Calibration
Methods.
8 LOCKING AND UNLOCKING THE CALIBRATION
To set a calibration lock PIN code:
1
Press the Menu softkey and select
Setup/Calibration Lock.
2
Set the four digit PIN Code using the
é
softkeys to select 0 to 9
ê
and the è softkey to move from the
first to the fourth digit .
3
Press Ok, Escape to cancel or Adjust
to amend the PIN code.
The è softkey will appear when the " X " changes to a number.
en-5
R
90°
www.elcometer.com

Publicidad

Tabla de contenido
loading

Tabla de contenido