Understanding Out-Of-Range Control Results - Accu-Chek Nano Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Control Tests

Understanding Out-of-Range Control Results

WARNING
The control range applies only to a control result. It only indicates that the test strips and meter are working
properly. Do not use a control result to interpret blood glucose results.
If the control result is out of range, do not use the meter until you solve the problem. Check this list to help
solve the problem.
Troubleshooting Checks
1. Were the test strips or control solutions
expired?
2. Did you wipe the tip of the control solution
bottle before use?
3. Were the caps on the test strip container and
the control solution bottle always closed
tightly?
4. Was the test strip used immediately after it
was removed from the test strip container?
5. Were the test strips and control solutions
stored in a cool, dry place?
6. Did you follow the directions?
7. Are you still unsure of the problem?
12
Action
Discard the test strips or control solution if either is
past the Use By date. If the control solution was
opened more than 3 months ago, discard it. Repeat
the control test with an unexpired test strip and an
unexpired control solution.
Wipe the tip of the bottle with a tissue. Repeat the
control test with a new test strip and a fresh drop of
control solution.
Replace the test strips or control solution if you think
either was uncapped for some time. Repeat the
control test.
Repeat the control test with a new test strip and a
fresh drop of control solution.
Repeat the control test with a properly stored test
strip or control solution.
Read Chapter 2, Control Tests, and repeat the control
test.
Contact the Accu-Chek Customer Care Service Center
at 1-800-858-8072.