Laboratory pump Laboport® N96
Prerequisites
Spare parts/tool Spare part/tool
Information on
the procedure
Translation of Original Operating Instruction, english, KNF
323241-323242 06/20
10.3 Replacing diaphragm and valve plates
à
Disconnect the pump from mains and ensure that it is volt-
age-free.
à
Clean the pump and free the pump of hazardous materi-
als.
à
Remove the hoses from the pneumatic pump inlet and
pump outlet.
à
If necessary, remove attachment parts such as silencer,
suction filter, etc., from the gas connections.
Spare part set*
Torx TX10 screwdriver with
torque indicator
Tab.13: Spare parts/tool
*acc. to Chapter 11 Spare parts and accessories
Diaphragm and valve plates/seals are the only wear parts in
the pumps. They are easy to replace.
Valve plates/seals and diaphragm should generally be re-
placed at the same time. If the diaphragm is not replaced at
the same time as the valve plates/seals, the specified output
of the pump can no longer be ensured after the maintenance
is performed.
Quantity
1
1