ELECTROMAGNETIC COMPATIBILITY
The units with PROCOM2 or C2020 automate needs special precautions regarding EMC.
The units with PROCOM2 or C2020 automate has to be installed and put into service according to the EMC requirements joined in the following tables
Guida e a d a ufa tu e 's de la atio – electromagnetic emissions – for all ME equipment and ME systems
The units with PROCOM2 or C2020 automate is intended for use in the electromagnetic environment specified below. The customer or the user of the units with
PROCOM2 or C2020 automate should assure that it is used in such an electromagnetic environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guida e a d a ufa tu e 's de la atio – electromagnetic immunity - for all ME equipment and ME systems
The units with PROCOM2 or C2020 automate is intended for use in the electromagnetic environment specified below. The customer or the user of the units with
PROCOM2 or C2020 automate should assure that it is used in such an electromagnetic environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
<5 % UT (>95 % dip in UT) for 0,5
Voltage dips, short
cycle
interruptions and voltage
40 % UT (60 % dip in UT) for 5
variations on power
cycles
supply input lines
70 % UT (30 % dip in UT) for 25
IEC 61000-4-11
cycles
<5 % UT (>95 % dip in UT) for 5 s
Power frequency (50/60
Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
56
Compliance
Group 1
Class B
Class A
Complies
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power supply lines
2 kV for power supply lines
±1 kV for input/output lines
±1 kV for input/output lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
3 A/m
Electromagnetic environment – guidance
The units with PROCOM2or C2020 automate uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The units with PROCOM2 or C2020 automate is suitable for use in all establishments,
including domestic establishments and those directly connected to the public low
voltage power supply network that supplies buildings used for domestic purposes
Compliance level
For 230 Vac / 50Hz
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
±6 kV contact
at least 30 %.
±8 kV air
For 230 Vac / 50Hz
Mains power quality should be that of a typical commercial or
hospital environment.
For 230 Vac / 50Hz
Mains power quality should be that of a typical commercial or
±1 kV mode differential
hospital environment.
±2 kV mode common
Mains power quality should be that of a typical commercial or
Pour 230 Vac / 50 Hz
hospital environment. If the user of the units with PROCOM2 or
0 Vac during 10 ms
C2020 automate requires continued operation during power
92 Vac during 100 ms
mains interruptions, it is recommended that the emergency
161 Vac during 1 s
mobile system for air, O2 and vacuum networks with PROCOM2
0 Vac during 5 s
automate be powered from an uninterruptible power supply or
a battery.
Power frequency magnetic fields should be at levels characteristic
of a typical location in a typical commercial or hospital
3 A/m for 50Hz
environment.
No components sensible in product to magnetic field.
Electromagnetic environment – guidance