Section IV — Initial Setup

Attaching the A/C Power Adapter

Plug the A/C Power Adapter (Part Number 383491) into the A/C adapter port
on the back of the pump. See Figure 1A to locate port on pump.
Note: It is important to make sure the power plug is pushed all the way
into the pump for proper operation. If the display screen shows the pump is
running on battery power after plugging in the A/C adaptor, check to make
sure the A/C adaptor plug is flush with the back of the pump.
Do not detach the connector by pulling on the cord only.

Battery Setup

The battery is packed partially charged. Charge battery 6 hours prior to use
off A/C power.
It is also recommended that the battery charge be "topped off" before battery-only operation is utilized. The battery pack will charge
continuously whenever the pump is plugged into a wall outlet, including during normal use of the pump with A/C power. 6 hours of
charging is required to fully recharge the battery pack.
A new battery pack will provide approximately 18 hours of service before needing to be recharged.

Attaching Pole Clamp

The Kangaroo™ Joey enteral feed and flush pump with pole clamp can be
attached to a vertical pole using the pole clamp which is included with the
pump. The pole clamp provides the ability to retain the A/C power adapter
to prevent accidental cord removal or cord loss. Simply wrap the cord
around the cord wrap posts to ensure that the cord stays with the pump.
To attach the pole clamp to the pump, simply slide the clamp onto the
mounting stud on the back of the pump (Figure 3A). When engaged,
turn the lever to secure attachment. Avoid overtightening or loosening
of handle.
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
A/C Adapter Port
Figure 2. Alignment of A/C Adapter plug with pump's
adapter port.
Figure 3A.
Figure 3B. Attaching and using the pole clamps.