Directions for Cleaning Rotor Assembly
Refer to General Cleaning Directions before starting.
Open the blue door enclosing the pump set loading area.
Loosen rotor set screw with 5/64" (2 mm) allen wrench and gently pull rotor off shaft. After removing rotor, avoid getting any
moisture in the rotor shaft opening.
Clean rollers thoroughly with warm soapy water, or isopropyl alcohol if necessary.
Be sure all parts of rotor are completely dry before putting it back onto shaft.
To replace rotor, align set screw on hub of rotor with the flattened section of the output shaft.
Push rotor into place and tighten set screw (Do not over tighten).

Preventative Maintenance

This pump must be periodically tested to assure proper functioning and safety. The recommended service interval is every two years.
Testing may be done at the user's Biomedical Engineering Department, an outside service, or by Covidien Factory Service. To arrange
for Covidien Factory Service in the US, call 1-800-962-9888. In Canada, call 1-877-664-8926. Outside of the US and Canada, please
contact your local Customer Service for more information.
If a pump malfunctions, please contact your Covidien Representative or call Customer Service for instruction.