Figure 1B. Kangaroo™ Joey enteral feed and flush pump with pole clamp set loading area.
Section II — Safety and Warnings
Note to healthcare personnel who provide training to lay operators or lay responsible organizations:
Be sure to include all of the Warnings below when providing training to lay operators, especially in a Home Care Environment.
Lay users should be instructed to contact Customer Service if there is a change in the performance of the pump. Additionally,
Lay Operators should be instructed on proper cleaning procedures to avoid hazards such as electric shock. Lay users should
also be trained on inappropriate environments for use (e.g., bathtub) of the pump. For guidance on training, please contact
Customer Service.
Caution: U.S. federal law restricts the sale of this device to physicians or to their direct representatives.
1. Read this booklet thoroughly before using the Kangaroo™ Joey enteral feed and flush pump with pole clamp.
2. Do not use this device near flammable anesthetics.
3. Disconnect power supply before cleaning or servicing.
4. The Kangaroo™ Joey enteral feed and flush pump with pole clamp utilizes a two prong, medical grade power supply cord
specifically designed for use only with the Kangaroo™ Joey feeding pump's Lithium Ion battery charging circuit. Use of an
alternate consumer style AC power adapter or DC car adapter may cause damage to the charging circuit and
battery of the Kangaroo™ Joey feeding pump. Use only the supplied power adapter to charge your Kangaroo™ Joey
feeding pump from an AC power source. Use only Kangaroo™ Joey enteral feed and flush pump with pole clamp A/C power
adapter (with built-in A/C adapter) with pump. See Section XIII - Service Part Numbers for replacement of A/C power adapter.
5. Danger: Use only Kangaroo™ Joey enteral feed or feed with flush pump sets with this device. Pump is not compatible with
other pump sets. Use of other feeding sets with this pump can create hazardous situations, including free-flow conditions that
can result in overfeeding, underfeeding, formula in the lungs, and death to a patient.
6. For performance tests, see Section VI - Performance Tests. For other integrity checks, consult with a qualified biomedical
technician or contact the manufacturer. (Section XI - Customer Service).
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
Feeding Set
Loading Area
3
Table of Contents