CLEANING
Clean the M-LNCS to LNC MAC Adapter by wiping it with a 70% isopropyl alcohol pad and allow it to dry.
CAUTION
• Do not soak or immerse the cable in any liquid solution. Do not attempt to sterilize.
SPECIFICATIONS
The M-LNCS to LNC MAC Adapter Cable is for use with the following sensors:
Sensors
M-LNCS Adtx/Adtx3
M-LNCS Pdtx/Pdtx3
M-LNCS Inf/Inf-3
M-LNCS Neo/Neo-3
M-LNCS NeoPt/NeoPt-3
M-LNCS DCI
M-LNCS DCIP
M-LNCS TC-I
NOTE: A
accuracy is a statistical calculation of the difference between device measurements and reference measurements.
rms
Approximately two-thirds of the device measurements fell within ± A
The saturation accuracy of the neonatal sensors were validated on adult volunteers and 1% was added to account for the
properties of fetal hemoglobin.
ENVIRONMENTAL
Operating Temperature
Storage Temperature
Relative Humidity
WARRANTY
Masimo warrants to the initial buyer only that this product, when used in accordance with the directions provided with the
Products by Masimo, will be free of defects in materials and workmanship for a period of six (6) months. Single use products
are warranted for single patient use only.
THE FOREGOING IS THE SOLE AND EXCLUSIVE WARRANTY APPLICABLE TO THE PRODUCTS SOLD BY MASIMO TO BUYER.
MASIMO EXPRESSLY DISCLAIMS ALL OTHER ORAL, EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION
ANY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE. MASIMO'S SOLE OBLIGATION AND
BUYER'S EXCLUSIVE REMEDY FOR BREACH OF ANY WARRANTY SHALL BE, AT MASIMO'S OPTION, TO REPAIR OR REPLACE
THE PRODUCT.

WARRANTY EXCLUSIONS

This warranty does not extend to any product that has been used in violation of the operating instructions supplied with the
product, or has been subject to misuse, neglect, accident or externally created damage. This warranty does not extend to any
product that has been connected to any unintended instrument or system, has been modified, or has been disassembled or
reassembled. This warranty does not extend to sensors or patient cables that have been reprocessed, reconditioned or recycled.
IN NO EVENT SHALL MASIMO BE LIABLE TO BUYER OR ANY OTHER PERSON FOR ANY INCIDENTAL, INDIRECT, SPECIAL OR
CONSEQUENTIAL DAMAGES (INCLUDING WITHOUT LIMITATION LOST PROFITS), EVEN IF ADVISED OF THE POSSIBILITY
THEREOF. IN NO EVENT SHALL MASIMO'S LIABILITY ARISING FROM ANY PRODUCTS SOLD TO BUYER (UNDER A CONTRACT,
WARRANTY, TORT OR OTHER CLAIM) EXCEED THE AMOUNT PAID BY BUYER FOR THE LOT OF PRODUCT(S) INVOLVED IN
SUCH CLAIM. IN NO EVENT SHALL MASIMO BE LIABLE FOR ANY DAMAGES ASSOCIATED WITH A PRODUCT THAT HAS BEEN
REPROCESSED, RECONDITIONED OR RECYCLED. THE LIMITATIONS IN THIS SECTION SHALL NOT BE DEEMED TO PRECLUDE
ANY LIABILITY THAT, UNDER APPLICABLE PRODUCTS LIABILITY LAW, CANNOT LEGALLY BE PRECLUDED BY CONTRACT.
Body Weight
> 30 kg
10–50 kg
3–20 kg
< 3 kg
> 40 kg
< 1 kg
> 30 kg
10–50 kg
> 30 kg
41°F to 104°F (5°C to 40°C)
-40°F to 158°F (-40°C to 70°C)
Up to 95%, non-condensing
Saturation Accuracy
(70–100% SpO
)
2
No Motion
2%
2%
2%
3%
2%
3%
2%
2%
3.5%
of the reference measurements in a controlled study.
rms
4
Pulse Rate Accuracy
(25–240 bpm)
No Motion
3 bpm
3 bpm
3 bpm
3 bpm
3 bpm
3 bpm
3 bpm
3 bpm
3 bpm
5583D-eIFU-0319