Con2:
Head
Head
Head
Head
Weig
Weig
Weig
Weig
er0
er1
er2
er3
ht1
ht2
ht3
Header0=02H
Header1 follow decimal point
Decimal point=0, header1=22H
Decimal point=1, header1=23H
Decimal point=2, header1=24H
Decimal point=3, header1=25H
Decimal point=4, header1=26H
Header2 follow weigh status, default value=20H
If in net mode (tare value not 0), header2=header2|01H
If gross weight "-", header2=header2|02H
If overload or gross weight "-", header2=header2|04H
If unstable, header2=header2|08H
If weighing unit=kg, header2=header2|10H
Header3 follow weighing unit
If weighing unit=g, header3=21H
If weighing unit=oz, header3=23H
Weight1~weight6: weighing data
Tare1~tare6: tare value
Terminator1: 0DH
Terminator2: 0AH
:
Con3
Header
Header
Weight
Weight
Weight
Weight
0
1
1
2
3
4
HEADER0=01H
Header1 follow weight "+" or "-"
When weight "+", header1="+", when weight "-", header="-"
Weight1~weight7: weight data (include decimal point)
Unit1~unit2: weight unit
Status: when stable, status=0, when unstable, status=1
Terminator1: 0DH
Terminator2: 0AH
BACKLIGHT
At startup, the backlight will automatically turn on
when the scale is in use. Enter this menu to change the
backlight setting.
1.
Press and hold the ZERO key
SET bF
will be shown on the display.
2. Press the ZERO key
PAGE 9 OF 36
OPERATION CONTINUED
Weig
Weig
Tare1 Tare2 Tare3 Tare4 Tare5 Tare6
ht4
ht5
ht6
Weight
Weight
Weight
Unit1
Unit2
Status
5
6
7
for 3 seconds.
to enter the backlight menu.
LEGAL FOR TRADE
Termin
Termin
When the indicator is used in trade or a legally
ator1
ator2
controlled application, it must be set up, verified and
sealed in accordance with local weights and measures
regulations. It is the responsibility of the purchaser to
ensure that all pertinent legal requirements are met.
CALIBRATION
1.
Break the seal on the indicator housing.
2. Open the housing.
3. See calibration process on page 5. Complete steps 1-5.
4. After step 5, press the calibration switch on the
mainboard. (See Figure 10)
Figure 10
5. Complete remaining calibration steps on page 5.
Termin
Termin
ator1
ator2
VERIFICATION
The local weights and measures official or authorized
service agent must perform the verification procedure.
SEALING
The local weights and measures official or authorized
service agent must apply a security seal to prevent
tampering with the settings.
SETUP MENU
3. Press the TARE key
shown on the display.
AU
•
– Backlight automatically turns on when the
scale is in use.
on
•
– Backlight is always on.
oFF
•
– Backlight is never on.
4. Press the ZERO key
5. Press the UNIT key
Top
Calibration Switch
until the desired setting is
to select the new setting.
to exit the sub menu.
0621 IH-5836