Capacitance measurement
WARNING
on live circuits. Safely discharge capacitor before testing.
Insert the black connector lead (S) into the COM (black)
input terminal (I) and the red connector lead (S) into the V
(red) terminal (M).
Set the function switch (K) to the Ω/CAP position.
Press the MODE/VFD button (H) until the nF symbol
appears on the display (F).
Connect the test probes (S) to the capacitor being tested.
If OL appears on the display (F), remove and discharge the
capacitor.
Read the capacitance on the display (F).
NOTICE
measurement, it can take several minutes before the final
reading stabilizes.
NOTICE
polarity (+, –) of its leads to ensure a precise measurement.
Frequency/duty cycle measurement
Insert the black connector lead (S) into the COM (black)
input terminal (I) and the red connector lead (S) into the V
(red) terminal (M).
Set the function switch (K) to the
22
Risk of electric shock! Never test capacitors
For very large values of capacitance
When measuring electrolytic capacitors, follow the
/ Hz / % position.