The approved entity parameters for the Canadian approval for the ACM500 module are listed in
Table 2-5.
64061976
|
08
|
11/2019
Table 2-5: Entity Values - Canada
Entity Values for Canadian Approval
U
10 VDC
i
I
300 mA
i
P
500 mW
i
C
120 nF
i
L
0 mH
i
METTLER TOLEDO ACM500 Installation Manual