3.2.
AC Power Considerations
No internal wiring is required for AC power, since the ACM500 is supplied with an integral power
cord. When installing the ACM500, make sure the unit is near a properly grounded socket-outlet
and that the outlet is easily accessible.
Make sure that the AC power source is within the operating range of the ACM500, 100 to 240 VAC.
3.3.
Mounting the Enclosure
The ACM500 enclosure is designed to mount to a flat, vertical surface such as an instrument panel
or industrial enclosure or wall. Observe location and environment considerations as described in
Chapter 1 of this manual, Introduction.
To mount the enclosure, follow these steps:
1. Mark the position of the mounting holes on the vertical surface using the dimensions shown in
2. The hardware to mount the enclosure to the vertical surface is not included with the terminal – it
64061976
|
08
|
11/2019
Figure 3-3, or by holding the enclosure up to the surface and marking through the bracket
holes on the sides of the enclosure. Measurements are provided in mm and inches.
must be supplied locally. Ensure that the mounting hardware is capable of supporting the
weight of the terminal, which is approximately 6 kg (13 lb). Using the locally-supplied
hardware, mount the ACM500 to the vertical surface.
Figure 3-3: Mounting Hole Pattern
METTLER TOLEDO ACM500 Installation Manual
3-3