steps:
• Press down the 'Memory/Save' button (18) for 3
seconds.
• The M1-M4 buttons will first light up blue and then
switch to white again.
• The LCD backlighting system will flash white once
after several seconds.
• Immediately press down the synchro button (17)
and keep it pressed down. The control systems
power down.
• If the entire backlighting system flashes white twice,
you can release the Synchro button (17).
The internal meter readings for the individual motors
have now been reset to '0' and the stored memory po-
sitions are correct again.
Use of the emergency lowering button consumes a
great deal of battery power, which means you will need
to replace the batteries.
The Swissflex Box _95RF does not work despite be-
ing switched on and all components being connected
properly
Disconnect the mains supply unit from the mains out-
let. Wait a minute and then reconnect it. If necessary,
connect a bedside lamp to check that mains voltage
(240 V) is present. If there is no current, check the fuse.
Press the "all motors down" button (7) to reset all
adjustments. The control of the Swissflex Box _95RF
contains microprocessors, so irregular power supply
can lead to malfunctions.
The radio remote control does not work (The back-
Problem
Possible cause
Remote control
No mains supply.
or control system
Manual switch or control system
not working.
defective.
Control
Thermoswitch on the transformer
systems sud-
may have been triggered or may be
denly no longer
defective.
travel / move.
Thermal cut-out in the transformer
may have been triggered or may be
defective.
Internal fuse may have been triggered
or may be defective.
No mains supply.
Supply line (mains and / or manual
switch / auxiliary control system) inter-
rupted.
The battery-
Block battery / block batteries flat.
operated electric
reset function
Block battery/batteries not connected.
cannot be acti-
vated.
24
Operating instructions
ground light does not work)
Insert new batteries. Where applicable, remove any
obstacles between the radio remote control and the
motor control.
After a power failure once power has been restored
to the system
To restart the initial setup of the Swissflex Box _95RF,
press button (30) on the switching power supply (28)
until all of the motors have powered down and this is
confirmed by the motor's emergency light blinking twi-
ce (can take 20 seconds).
Technical data
Supply voltage: 110-240 V
Remote control power supply: 3 x AAA Batteries
Power rating motors:
Intermittent operation FROM 2 min / 18 min
Power rating user socket : 2000 W / 10 A
Thrust motors: 2 x 6000 N / 2 x 4000 N
Power supply for emergency lowering:
2 x 9 V block battery
Solution
Restore mains connection.
Contact your supplier / stockist.
Leave the control system in the rest position for 20-30
minutes.
Contact your supplier / stockist.
Contact your supplier / stockist.
Restore mains connection.
Check the supply line and restore contact if applicable.
Check and replace the block battery / block batteries if
necessary.
Connect the block battery / block batteries.
EN