Cleaning Of Accessories; Maintenance - Ca-Mi New Vapinal Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

GB
Clean the accessories as described below.
To clean the steam conveying nebulizer (1):
Remove the steam conveying nebulizer from its seat in the sprayer nozzle by loosening the screw.
Submerge the nebulizer into a common liquid descaler, soaking it there until all hard water deposits have
disappeared. Follow the manufacturer's instructions for the liquid descaler. Rinse thoroughly with running water.
To clean the sprayer nozzle.
Remove the steam conveying nebulizer from its seat in the sprayer nozzle (2) by loosening the screw (4).
Remove the sprayer guard and, using the CH17 wrench provided, unscrew the nut that holds the part on the
boiler. Hold the plastic body of the nozzle in your hand and unscrew counterclockwise.
To prevent scale deposits from forming, fill a regular glass with 50% water and 50% vinegar. Soak the nozzle for
one hour in the prepared solution, then rinse under running water.
To remove scale deposits on the sprayer, fill a regular glass with 50% water and 50% common descaler.
Soak the part for 12 hours in the prepared solution, then rinse well with running water.
You may also periodically use the pin (14) provided to remove deposits from the nozzle's central hole.
If the sprayer has been taken apart for cleaning, reassemble it correctly: the seal of the nut must be flush with the
socket (5) that sticks out from the side of the unit.
The NEW VAPINAL Thermal Water Inhaler does not need maintenance or lubrication.
It is necessary to check functioning and instrument before every use. Unpack the instrument and always check
integrity of plastic parts and feeding cable., they might have been damaged during previous use.
After having filled the boiler with water, screw the valve cap back on and check that the rubber gasket is flush with
the bushing of the boiler that sticks out form the device.
Then plug the cord into the mains. Turn on the switch.
Verify that there are no steam leaks from the valve cap and/or sprayer nut.
At the end of the inhalation treatment, let cool and remove the water from the boiler (see chapter
INSTRUCTIONS FOR USE). The device is protected with a protection fuse (1 x F4A L 250V) that can easily be
replaced by the operator. It is located inside the socket at the back of the device. To replace it, always make sure it
matches the reported type and value. Pull out the plug form the socket before replacing the fuse.
Problem
1.
The water is not being drawn
from the bottle.
2.
The device turns off mid-
treatment.
3.
Steam is not being released.
4.
The steam comes out with too
little pressure from the
nebulizer nozzle.
5.
The device turns off without
completely vaporizing the water
in the bottle.
6.
The device turns off during
inhalation treatment.
Problems 1 - 2 - 3 - 4 - 5 - 6
In case of malfunction, contact a qualified CA-MI technician and/or technician authorized by CA-MI.
Never open the device.
CLEANING THE ACCESSORIES

MAINTENANCE

Cause
Sprayer clogged with
limescale.
Sprayer clogged with
limescale.
Power cord not plugged in.
Safety thermostat has been
triggered.
Nebulizer nozzle clogged.
Nebulizer nozzle clogged.
Safety thermostat has been
triggered.
None of the solutions has
been effective.
20
Solution
Allow the unit to cool, remove the sprayer with
the wrench provided and clean as indicated in
the cleaning chapter.
Allow the unit to cool, remove the sprayer with
the wrench provided and clean as indicated in
the cleaning chapter.
Make sure that the cord of the device is
correctly plugged in to the mains and that the
switch is turned to position I.
Check that there is water in the boiler. If the
safety thermostat has been triggered, contact
CA-MI Technical Service.
Allow the unit to cool, remove the sprayer with
the wrench provided and clean as indicated in
the cleaning chapter.
The nebulizer nozzle may be blocked. For
cleaning, see the paragraph "Cleaning the
Accessories".
The safety thermostat could have been
triggered. Wait 30 minutes for the device to
cool. If the problem persists, contact CA-MI
Technical Service.
Contact your dealer or the CA-MI Customer
Service Center.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Re 420000

Tabla de contenido