4 Disinfect the mirror sleeve of the CEREC Omnicam with the HLD set
12
Operating instructions for CEREC Bluecam and CEREC Omnicam Care and HLD (high-level disinfection)
Disinfect the mirror sleeve
NOTICE
The plug does not seal the mirror sleeve watertight.
1. Press the green plug into the interior of the mirror sleeve in order to
ensure that no particle contamination or fluids penetrate the inside of
the mirror sleeve.
Disinfectants: ProCide-D Plus or CIDEX
manufacturers' safety indications!
2. Fill the HLD container using one of the following disinfectants for the
high-level disinfection: ProCide-D
disinfectant up to the "Omnicam" mark of 50ml.
NOTICE
Only place the mirror sleeve into the container and not the entire
camera.
Ensure that the mirror sleeve is placed upright in the disinfectant (with
the coated sapphire glass down) so that the solution cannot penetrate
the inside of the mirror sleeve.
CAUTION
Sirona Dental Systems GmbH
®
OPA – Observe the
®
®
Plus or CIDEX
OPA. Fill the
62 82 839 D3492
D3492.201.02.02.02 01.2016