Operating instructions for CEREC Bluecam and CEREC Omnicam Care and HLD (high-level disinfection)
62 82 839 D3492
D3492.201.02.02.02 01.2016
4 Disinfect the mirror sleeve of the CEREC Omnicam with the HLD set
3. Insert the mirror sleeve in the correct position (coated sapphire glass
down).
4. Place the lid on the container and leave it for as long as is specified
in the disinfectant manufacturer's documentation.
5. Remove the mirror sleeve from the container.
6. Rinse the exterior of the mirror sleeve. Observe also the directions for
using the disinfectant. Ensure that no water penetrates the mirror
sleeve.
The outlet window (coated sapphire glass) must be pointing down
when rinsing the mirror sleeve. The mirror sleeve may only be rinsed
below the slot (A) so that no water penetrates into the mirror sleeve.
7. Dry the mirror sleeve using a soft, lint-free cloth.