INDICATIONS FOR USE: VIOS
The Vios
is a tabletop, AC-powered air compressor intended to provide a source of compressed air for use with
jet nebulizers. The Vios
environments.
B. SAFETY PRECAUTIONS - VIOS
WARNING
Use only with nebulizers and tubing authorized by PARI
for this device. Other nebulizers and tubing may result in
improper treatment.
When using electrical and medical products, especially
when children are present, basic safety precautions must
be followed.
DANGER
To reduce the risk of serious or fatal injury from
electrocution:
1. Do not place or store product where it can fall or be
pulled into a bathtub, sink, water, or other liquid. Do
not place in or drop into water or other liquid. Do not
use while bathing.
2. Do not reach for a product that has fallen into water or
other liquid. Unplug immediately.
WARNING
To reduce the risk of serious or fatal injury from
electrocution, fi re or burns and to reduce the risk of
damage and malfunction to the product:
1. Before use, check the serial number label on the
bottom of the PARI Vios
the voltage and current indicated on the compressor
label correspond to the voltage and current available.
2. Never operate this product if it has a damaged cord
or plug, if it is not working properly, if it has been
dropped or damaged, or if it has been exposed to any
liquids inside the case. Return the product to PARI for
examination and repair.
3. Keep all electrical cords away from heated surfaces. Do
not overload wall outlets or use extension cords.
4. This appliance has a polarized plug (one blade is wider
than the other). As a safety feature, this plug will fi t in
a polarized outlet only one way. If the plug does not fi t
fully in the outlet, reverse the plug.
5. Never block the air vents of this compressor or place it
on a soft surface, such as a bed or couch, where the air
vents may be blocked. Keep the air vents free of lint,
hair, and similar substances. Blocked vents could cause
excessive heat buildup. This product should never be
left unattended when plugged in.
Vios IFU- Sprint EN 0029.indd 3
Vios IFU- Sprint EN 0029.indd 3
TM
AEROSOL DELIVERY SYSTEM
Compressor is intended for adult and pediatric patients for use in hospital, clinic, or home
TM
AEROSOL DELIVERY SYSTEM
Compressor to ensure that
WARNING
6. Do not operate outdoors.
7. Not suitable for use in the presence of fl ammable
anesthetic mixture with air or oxygen or nitrous oxide.
8. Do not operate where oxygen is being administered in
a closed environment such as an oxygen tent.
9. Never drop or insert any object into any openings.
10. Always unplug this product immediately after using.
11. Be sure the PARI Vios
to cleaning.
12. Do not open or disassemble the PARI Vios
Compressor. Refer all servicing to PARI.
WARNING
To reduce risk that the compressor will fail to operate
properly as a result of decreased electromagnetic
interference, resulting in improper treatment:
1. This device needs special precautions regarding
electromagnetic compatibility ("EMC") and must be
installed and put into service according to the EMC
information provided in these Instructions For Use.
2. Portable and mobile radio frequency ("RF")
communication devices can disrupt electrical medical
equipment. Technical EMC information is available on
pages 9-10.
3. Use only with accessories and cables authorized
by PARI for this device. Use of this device with
unauthorized accessories and cables may result in
increased electromagnetic emissions or decreased
electromagnetic immunity of the device.
4. This device should not be used adjacent to, or stacked
with other equipment. If adjacent or stacked use is
necessary, this device should be observed to verify
normal operation.
Questions? 1.800.FAST.NEB (327.8632) | PARI.com/VIOS
Compressor is unplugged prior
EN-2
6/16/10 11:47:09 PM
6/16/10 11:47:09 PM