Surgery With Saline Solution (Nacl); Preparations On The Unit; Presetting - Sirona C1+ Instrucciones De Uso

Ocultar thumbs Ver también para C1+:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Sirona Dental Systems GmbH
19

Surgery with saline solution (NaCl)

19.1

Preparations on the unit

19.2

Presetting

C
Ident
1
59 06 446 D 3386
D 3386.201.01.08.02
C
19 Surgery with saline solution (NaCl)
Hook NaCl bottle (1) in place.
Attach the peristaltic pump set (2).
Push short end of hose (3) with cannula as shown
through stopper and into NaCl bottle.
The regulator in the hose clip (4) must be in the top
position (completely open).
Run the long hose (5) along the corresponding mo-
tor hose up to the right-angle handpiece and fasten
with the clips (6).
Fit coupling (7) onto hose and connect it with the thin
silicone hose (8).
Connect spray clip (9) with the thin hose and attach
to angle piece.
The peristaltic pump set (2) is a disposable article and
can be purchased as consumable material under the
article number F 58707 directly from the manufacturer in
packs of 10 pcs.
Ordering address:
Satelec
Industriestrasse 9,
D-40822 Mettmann
CAUTION
The flow of the pump must be at least 70 ml/min for rea-
sons of safe cooling and to prevent a pressure rise in the
hose (risk of bursting). This is not always the case when
non-Sirona ultrasound tips are used.
If corresponding NaCl pumps are fitted to the dentist ele-
ment, the possibility of selecting saline solution NaCl for
surgery can be preset for a maximum of 2 burr instru-
ment receptacles (including SIROSONIC L).
For this purpose actuate the
dialog.
The Service dialog is displayed.
In the Service dialog, actuate the
19.1 Preparations on the unit
key in the assistant
1
key.
75

Publicidad

Tabla de contenido
loading

Tabla de contenido