Siemens XRS-5 Guia De Inicio Rapido página 9

Ocultar thumbs Ver también para XRS-5:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 43
Mounting
Mount the transducer so that it is above the maximum material level by more than
the blanking value for accurate results. Refer to the associated transceiver manual
for information on setting the blanking value.
The transducer must be mounted so that the axis of transmission is perpendicular to
the liquid's surface.
Do not over tighten mounting. Hand tightening of the mounting hardware is
sufficient.
Consider the optional temperature sensor when mounting the transducer.
Suspended Conduit
rigid metal
conduit
coupling
Flexible conduit transducer should not be
subjected to wind, vibration or jarring.
Submersible
rigid metal
conduit
coupling
submergence
shield
Submersible transducer, used in
applications where flooding is possible.
A5E32299685
XRS-5 Series Transducers – QUICK START MANUAL
transducer
transducer
Plywood mounting provides excellent
isolation, but must be rigid enough to avoid
flexing if subjected to loading.
Bracket
flexible conduit
steel channel
coupling
Plywood
coupling
Page EN-7

Publicidad

Tabla de contenido
loading

Tabla de contenido