TECHNICAL DETAILS

AC INPUT RATING:
FUSE RATING:
LAMP RATING:
Protection from Electric Shock:
Protection against ingress of water:
Operation Mode:
Operating Environment:
Temperature:
Relative Humidity:
Atmospheric Pressure:
Transport and Storage Conditions:
Temperature:
Relative Humidity:
Atmospheric Pressure:
Dimensions:
Height: (base):
4.25 inches
(with hand-piece):
8.50 inches
Width:
5.30 inches
Length:
6.80 inches
NOTE: - Models 700, 701, 702 and 703 meet all of the following Safety Standards.
- Model 704 is pending approval certifications.
Complies to following Directives/Standards:
93/42/EEC
- MDD
IEC 601-1
- 1988 - Medical Electrical Equipment (General Requirements for safety)
IEC 601-1 Am.1
- 1991 - Amendment 1 to IEC 601-1
IEC 601-1 Am.2
- 1995 - Amendment 2 to IEC 601-1
EN 60601-1
- 1993 - Medical Electrical Equipment (General Requirements for safety)
EN 60601-1-2
- 1993 - IEC 601-1-2: Medical electrical equipment Part 1: General requirements for safety 2:
CSA C22.2 no.601.1
- M90-Medical Electrical Equipment Part 1: General Requirements for Safety
UL 2601-1
- 1994 - Medical Electrical Equipment: Part 1: General Requirements for Safety.
Model 700 also complies with EMC requirements for FDA 510(k) premarket notification submissions.
NOTE: A Spectrum ® 800 Curing Light Repair Manual is available upon request from DENTSPLY.
MODEL 700:
MODEL 701, 702, 703:
MODEL 704:
MODEL 700, 704:
MODEL 701, 702, 703:
75W, 12V
Type BF Applied Part
IEC Equipment Class I
(Ordinary Equipment) IXPO
Continuous operation with intermittent loading:
(2 min./5 min. = 2 min. ON / 5 min. OFF)
NOTE: Equipment not suitable for use in the presence of a flammable
anesthetic mixture with air or with nitrous oxide.
10 to 25 deg. Celsius
30% to 75% (non-condensing)
700 to 1060 hPa
-10 to +60 deg. Celsius
10% to 100% (non-condensing)
500 to 1060 hPa
( 11.0 cm)
( 22.0 cm)
( 13.5 cm)
( 17.5 cm)
Directive
Collateral standard: Electromagnetic compatibility - Requirements and tests
(Amendment 1 and 2)
120V, 50/60Hz, 2.0A
230V, 50/60Hz, 1.0A
100V, 50/60Hz, 2.0A
T2.5A, 250V
T1.6A, 250V
Hand-piece cord
length (approx.):
80.0 inches
Power cord length:
60 - 120 inches (152-305cm)
Weight: 5.2 pounds
5
English
(203.0 cm)
(2.36 kg)