INTRODUCTION
The Yellow Jacket
52030F Thermal Imager is an accurate, high-quality
TM
handheld thermal imaging camera. The 52030F has been designed
exclusively for the purpose of performing elevated skin temperature
screening. The 52030F thermal imager meets the recommendations
established by the U.S Food & Drug Administration (FDA) for elevated skin
temperature screening. The 52030F has a measurement accuracy of ±0.9F
(±0.5C) over a temperature screening range of 93.2-102.2°F (34-39°C).
The thermal imaging camera has been calibrated in the factory using
a blackbody temperature reference source. No on-site calibration is
required.
Meets international performance standards for thermographic devices.
Not FDA-cleared or approved.

KEY FEATURES

• Designed exclusively for elevated skin temperature screening
• 80x60 thermal resolution or 4800 temperature measurement points
• 9Hz frame rate. The camera creates images at the rate of 9 per second
• Hot, Cold, Center temperature tracking
• Large, easy-to-read graphical TFT display
• Buzzer alarm
• 6 hours operating time with rechargeable battery
• Compact, rugged design
PRODUCT USE RECOMMENDATIONS
• Please read this Operating Manual carefully before using the
thermal imager
• The optimal measuring distance for this product is 3 feet.
• To ensure measurement accuracy, please use the product in an
operating environment of 59°F to 86°F (15°C to 30°C), <85% relative
humidity (non-condensing).
• Please use the thermal imager indoors and in a room with minimal air
movement. When changing to a new environment, please turn on the
thermal imager 10 to 15 minutes before you begin taking measurements.
• The ambient temperature for measuring must be stable. Do not measure
in places where there is a lot of airflow from fans or air outlets.
• If the person being measured has been in a place where the temperature
was significantly warmer or cooler than the measurement environment,
have that person remain in the measurement room for 10 to 30 minutes
before taking their temperature.
• The 52030F measures the surface temperature of an object. If
temperature compensation is needed, please adjust in the Settings menu.
• The thermal imager is self-calibrating. If the reading jumps quickly, allow
the measurement to steady before taking a reading.
• After measuring extremely high or low temperature objects, leave the
product alone for about 10 minutes before next use.
• Do not use the product in places with strong sunlight or electromagnetic
interference.
• Please do not use this product in flammable, explosive, steamy, wet or
corrosive environments.
• Please stop using the product if it has been damaged or modified to avoid
inaccurate measurement results.
• Highly Recommended: For high measurement accuracy, ensure that the
emissivity setting in the thermal imager matches the surface material (ex.
human skin) you are measuring.
• For better results, , please warm-up the thermal imager for 10 minutes
before performing measurements.
• When the battery is being charged, the internal temperature of the
product will increase. This can impact the accuracy of temperature
measurements. Therefore, it is not recommended to take measurements
while the product is being charged or right after it has been charged.
IMPORTANT INFORMATION CONCERNING
ELEVATED SKIN TEMPERATURE SCREENING
• Temperature measurement should not be solely or primarily relied upon
to diagnose or exclude a diagnosis of COVID-19, or any other disease