Technical Data - DPM Dew Point Meter Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Conforms to: ISO 8502-4, BS7079-B and ASTM D3276
Operating Range: -40° C to +80° C
PosiTector DPM, DPM S, DPM A, DPM D, DPM IR
Surface
Temp.
IR Surface
Temp.**
Air
Temp.
Humidity
Wind Speed*
*DPM A Models Only
**DPM IR Models Only
***At Ta = 23° C
PosiTector DPM IR:
- Distance to Target Size Ratio (D:S) – 4:1
- Response Time: <500 milliseconds (95% response)
- Spectral Response: 5.5 - 14 µm

Technical Data

Technical Data
Range
-40° to 80° C
80° to 190° C
-40° to 175° F
175° to 375° F
-70° to 380° C
-94° to 716° F
-40° to 80° C
-40° to 175° F
0 to 100%
0 - 20 m/s
0 - 65 ft/s
0 - 3900 fpm,
0 - 72 km/h
0 - 44 mph
0 - 38 kn
12
-40° F to +175° F
Accuracy
Resolution
±0.5° C
±1.5° C
±1° F
±3° F
±1° C +
(0.01°C/°C)***
±2° F +
(0.01°F/°F)***
±0.5° C
±1° F
±3 %
±3%
Full Scale
+ least
significant
digit
0.1° C
0.1° C
0.1° F
0.1° F
0.1° C
0.1° F
0.1° C
0.1° F
0.1 %
0.1 m/s
0.1 ft/s
1 fpm,
0.1 km/h
0.1 mph
0.1 kn

Publicidad

Tabla de contenido
loading

Tabla de contenido