Processing Ipas MVA Aspirators and Adapters
Method#
Chlorine†
High-Level
Dilute to 0.5%.
Disinfection
(HLD)
Boiling water*
(Ipas MVA Plus aspirator
and adapters only)
2% Glutaraldehyde
(Cidex)†
Follow manufacturer's
instructions for mixing.
Glutaraldehyde (other
solutions)†
Follow manufacturer's
instructions for mixing.
Steam autoclave*
Sterilize
(Ipas MVA Plus aspirator
and adapters only)
2% Glutaraldehyde
(Cidex)†
Follow manufacturer's
instructions for mixing.
Glutaraldehyde
(other solutions)†
Follow manufacturer's
instructions for
mixing.
STERRAD
100S
(MVA Plus and
adapters only)
# Ipas aspirators must be HIGH-LEVEL DISINFECTED OR STERILIZED BETWEEN USES.
* CAUTION: Never boil or steam autoclave the plungers from the Ipas Single-Valve aspirator as they will emit
formaldehyde. Do not interchange plungers between aspirator types.
† Glutaraldehyde and chlorine are hazardous substances. When processing instruments, take necessary precautions, such
as using personal protective equipment. Refer to the manufacturer's safety instructions to establish safe use.
Agent
Time
20 minutes
20 minutes
20 minutes
or follow
manufacturer's
instructions
Follow
manufacturer's
instructions
Sterility is achieved
at 121°C (250°F)
for 30 minutes with
pressure of 106 kPa
(15 lbs/in
2
Do not use other
autoclave settings.
Specifically, do not
use higher settings
for shorter periods
of time (known as
"flash autoclaving.")
10 hours
Follow
manufacturer's
instructions
55 minutes
Completely immerse disassembled parts. After processing,
rinse all parts with water. Discard solution daily or sooner if
solution becomes cloudy.
Disassembled parts do not need to be fully immersed.
Bring to room temperature before use.
Completely immerse disassembled parts. After processing,
rinse all parts with water. Discard solution 14 days after
mixing or sooner if solution becomes cloudy. Do not use
below 77° F (25° C).
Completely immerse disassembled parts. After processing,
rinse all parts with water. Usually discard solution 14 days
after mixing or sooner if solution becomes cloudy.
Place the disassembled aspirator in linen or paper.
Steam must penetrate all surfaces. Parts should not touch
and should be arranged so openings are not obstructed,
permitting drainage.
)
With the Ipas MVA Plus, the collar stop must be completely
removed (not held with the retaining clip).
Bring to room temperature before use.
Completely immerse disassembled parts. After processing,
rinse all parts with water. Discard solution 14 days after
mixing or sooner if solution becomes cloudy. Do not use
below 77° F (25° C).
Completely immerse disassembled parts. After processing,
rinse all parts with water. Usually discard solution 14 days
after mixing or sooner if solution becomes cloudy.
Place the disassembled aspirator along with a chemical
indicator strip in an approved tray or peel pack.

Precautions