General Specification; Trouble Shooting - SSCOR Quickdraw 2403 Serie Instrucciones De Operación Y Mantención

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 34
General Specifications
CHARACTERISTICS
Size
Weight
Vacuum Pump
Regulator
Battery
Collection Canister (13)
Patient tubing
(Type BF Applied Part)
Suction Tip
(Type BF Applied Part)

Trouble Shooting

PROBLEM
Does not function when
switch is turned on (DC
Power)
Battery Condition
indicator lights not lit
No suction when pump
is running
7
SPECIFICATIONS
12-1/2"L x 4"H x 4"W (32cm L x 11cm H x 11cm W)
2.6 lbs (1.18 kg)
12V DC, 0.7 A. Exceeds 500mmHg.
Lower levels of negative pressure will be observed at altitude.
Reduces vacuum from -500+mmHg (High Vacuum / Low Flow) to approximately
-85mmHg (Low Vacuum / Low Flow). Typically -70mmHg to -100mmHg but the range can
be larger due to conditions such as variances in battery charge and condition.
12V DC Sealed Lead Acid replacement part #80611-100
Alkaline battery replacement part #80613-100
10xAAA SSCOR Battery Holder part # 80615-100 (Batteries not supplied). FOR USE
WITH THE MODEL 2403 SERIES ONLY.
Please read page 6 for battery care suggestions
300cc/ml capacity. Non Sterile, Single Use
Model 2480: Standard replacement Quickdraw canister (shown on page 6)
Model 2488: Barbed Quickdraw canister (shown on front cover)
Vinyl tubing 9/32"ID x 36"L (7.1mm ID x 91.44 cm L)
SSCOR part #43203
HI-D® "Big Stick®" SSCOR part #44241
POSSIBLE CAUSE
Battery discharged
Molex connections disconnected
Damaged PC Board
Damaged PC Board
Canister not secure to chassis
Regulator vent is open
Tip cover is on the catheter
Fluids have shut down the filter in the
canister
CORRECTIVE ACTION
Install a new battery
Open unit and re-connect Molex connections
Replace PC Board
Replace PC Board
Latch canister to chassis
Replace cap on vent
Remove tip cover from catheter
Replace canister
Quickdraw - Model 2403 Series