4-1. Calibration Using the Internal Mass
(One-Touch Calibration, FZ-i / FZ-i WP / FZ-GD Series Only)
This function calibrates the balance using the internal mass.
1.
Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the weighing
pan.
2.
Place the accessory breeze break on the balance and press the
.
3.
The balance performs calibration using the internal mass. Do not allow vibration or drafts to affect the
balance.
4.
After the calibration, if the "GLP output (
outputs a sensitivity adjustment report.
5.
The balance will automatically return to weighing mode after calibration.
4. Calibration
)" parameter of the function table is set, the balance
7
key. The balance displays