Angle Calibration
Angle Calibration checks the alignment of the hub and disk. The results are analyzed and
reported as pass or fail. If the measurements pass the settings will be saved. If the analysis
fails, contact Support.
If performing the LED Calibration and the Angle Calibration, the LED Calibration should be
performed before the Angle Calibration.
1. Follow the Cleaning procedure on page 19 to clean the light chamber and optic lenses.
2. Tap
to go to settings.
3. Tap Calibration.
4. Tap Angle Cals.
5. Remove the Meter Check Disk (Code 1705/1705-EU) from the foil pouch. DO NOT
remove the black cover from the disk.
6. Insert the Meter Check Disk. Close the lid.
Start
7. Tap
to begin the meter check procedure.
8. If the meter is performing satisfactorily, "Angle Calibration Successful" will be
displayed and the settings will be saved. If the meter is not performing satisfactorily,
"Angle Calibration Unsuccessful. Contact Support" will be displayed.
9. Tap
to return to the Testing Menu and resume testing.
phone: 800-344-3100 option 3 | Mon-Fri 9am-5pm EST

HELPFUL HINTS

¬ Do not touch top or bottom of disk. Handle disk by the edge.
¬ Do not fill disk while in the meter. Fill disk on clean, dry surface.
¬ Fill the disk on a dark surface to more easily see the sample water.
¬ The disk should not contain any large air bubbles. Air bubbles will result in erroneous
results.
¬ Always use the disk cover.
¬ Only the Universal Disk Cover (Code 1719) can be used with the WaterLink
Touch
.
®
¬ Empty syringe of old sample before filling with next sample.
¬ Remove filled disk from meter after testing. Do not travel with filled disks in meter.
They may leak.
¬ Keep the chamber clean and dry. Gently swab LED and photodiode lenses located
around the hub with a cotton swab dampened with streak-free window cleaner. Do
not use alcohol. It will leave a thin film over the lenses when dry.
MAINTENANCE
CLEANING
The optical system of the WaterLink
performance. Dry the disk with a lint-free wipe before placing it into the chamber to avoid
introducing moisture. For best results, store the instrument in an area that is dry and free
from aggressive chemical vapors. Clean the exterior housing with a damp, lint-free cloth.
Do not allow water to enter the light chamber or any other parts of the meter. To clean the
light chamber and optic lenses, point a can of compressed air into the light chamber and
email: softwaresupport@lamotte.com
Spin Touch
®
17
must be kept clean and dry for optimal
®
Spin
®