¬ Do not place objects on the screen that will scratch or damage it.
¬ Avoid touching the screen with wet fingers.
TESTING
1. Press
and hold until the meter turns on.
2. Tap
. Select a Water Source type. Tap
3. Tap
. Select a disk series (found on disk packaging). Tap
NOTE: Disk Series are limited by Water Source Type selection.
4. Remove a disk from the packaging.
5. Use the syringe (1189) to fill the disk with the water sample.
6. Insert the disk. Cover the disk with the Universal Disk Cover (1719). Close the lid.
7. Tap
to start test. Tap
disk.
8. The results will be displayed.
9. Choose an option.
¬ Tap
¬ Tap the highlighted
not enabled.
¬ Tap the hightlighted
Printer.
¬ Tap the highlighted
¬ Tap
to return to the Test Screen.
10. Press and hold
NOTE: For water samples over 100 °F (38 °C) subtract 0.3 from pH result or, for the most accurate
result, wait until water sample is below 90 °F (32 °C) to test.
Remove residue daily. Residue will damage the meter and cause inaccurate results. See Cleaning, page
19.
Source
Treated Water
Disk Series
DW13
16-Apr-20
12:14 PM
to cancel the test. If the test is cancelled discard the
to add a tag.
to save the test results to the test log if Auto Save is
to send the results to the enabled Mobile Bluetooth
to send the results to a Bluetooth enabled device.
for 2 seconds to turn the meter off.
Tap to
Start Test
to confirm.
6
to confirm.