23. If the measured values are within the required tolerances, the Passes Test icon will be displayed.
Press the button adjacent to the Forwards Arrow icon, to continue, or press the button adjacent to
the Cross icon to abort the calibration process.
If the measured values are not within the required tolerances, the Fails Test icon is shown with an
error number. Refer to section 11 for a list of error codes.
Press the button adjacent to the Circular Arrow icon to repeat the calibration solution
measurement or press the button adjacent to the Back Arrow icon to return to the blank solution
measurement screen.
24. Following successful calibration, the instrument display will return to the micro volume accessory
settings screen and the new path length calibration factors will be shown.
21. The Calibration Solution Measurement icon will
be displayed. Open the instrument lid and use a lint
free cloth to clean the read heads. Open the
instrument lid and pipette a 2.0μl aliquot of the
calibration solution onto the lower read head. Close
the instrument lid and press the key adjacent to the
Tick
icon
to
measurement.
22. The unit takes two readings at 260 and 330nm, at
both 0.5mm and 0.2mm path lengths. The variation
from the expected Absorbance values are displayed.
initiate
the
calibration
solution
25