NOTE: If the pressure level is consistently low, check for any leakage (tubes or air hoses). If neces-
sary, replace any damaged tubes or hoses or contact your local qualified dealer for repair.

20. GUIDANCE AND MANUFACTURER'S DECLARATION

20.1 Guidance and Manufacturer's Declaration - Electromagnetic Emissions

This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
EMISSIONS TEST
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations /
Flicker emissions
IEC61000-3-3

20.2 Guidance and Manufacturer's Declaration - Electromagnetic Immunity

This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Immunity Test
Electrostatic Discharge
(ESD) IEC61000-4-2
Electrical fast
transient/ burst
IEC61000-4-4
Immunity Test
Surge
IEC61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC61000-4-11
Power frequency
(50/60Hz) magnetic field
IEC61000-4-8
COMPLIANCE
The device uses RF energy only for its internal function. Therefore, its
Group 1
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment
Class B
Class A
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
Complies
power supply network
IEC60601 test level
± 6kV contact
± 8kV air
± 2kV for power supply line
± 1kV for input/out line
IEC60601 test level
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5 % UT (>95 % dip
in UT)for 0,5 cycle
40 % UT (60 % dip
in UT)for 5 cycles
70 % UT (30 % dip
in UT)for 25 cycles
<5 % UT (>95 % dip
in UT)for 5 sec
3 A/m
NOTE:
UT is the A.C. mains voltage prior to the application of the test level
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
Compliance
Electromagnetic Environment - Guidance
± 6kV contact
± 8kV air
± 2kV for power
supply line
± 1kV for input/
atypical commercial or hospital environment
out line
Compliance
Electromagnetic Environment - Guidance
± 1 kV line(s) to line(s)
atypical commercial or hospital environment
<5 % UT (>95 % dip
in UT) for 0,5 cycle
Mains power quality should be that of atypical
40 % UT (60 % dip
commercial or hospital environment. If the
in UT) for 5 cycles
user of this device requires continued opera-
tion during power mains interruptions, it is re-
70 % UT (30 % dip
commended that the device be powered from
in UT) for 25 cycles
an uninterruptible power supply or a battery
<5 % UT (>95 % dip
in UT) for 5 sec
Power frequency magnetic fields should be
3 A/m
at levels characteristic of atypical location in
a typical commercial or hospital environment
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative hu-
midity should be at least 30 %.
Mains power quality should be that of
Mains power quality should be that of