Trouble Shooting Guide

Problem
Possible Cause
Rectal Light Handle is not
plugged into the power
supply.
Power Supply is not
plugged in.
Faulty Lamp
Lamp does not
light.
Loose Lamp
Wrong Lamp
Damaged Lamp
Rectal Light
Handle does not
Rectal Light Handle/
stay plugged
transformer interconnect
into the
plug is damaged.
transformer
cord.
Faulty Lamp
Light output
seems low.
Dirty lamp lens
This guide is intended to provide assistance towards determining the fault
condition of this product. Please contact the authorized Welch Allyn Service
Center listed within the product owners manual for additional assistance.
Solution
Verify that the Rectal Light Handle
is properly plugged into power
supply cord.
Verify the power supply is properly
plugged into the outlet.
Replace with a new lamp.
Verify that the lamp is secured into
the Rectal Light Handle.
Verify that the lamp is a REF
07800 Welch Allyn lamp.
Inspect the lamp and the o-ring for
damage. Replace with REF 07800
lamp if damage is found.
Inspect interconnect plug for
damage. Replace either or both
REF 73210 or REF 73305, 73322,
73324, 73326.
Replace with a new lamp.
Inspect lamp lens for debris and
clean.