Commissioning
3.3 Setting the Parameters
3. Press <Up/Down Arrow> keys to select a configuration that approximates the conditions
4. Press <ENTER> key to save selection. The [Anomaly Diams] menu will appear.
①
Use this menu cell to select the pipe configuration that most accurately represents the upstream pipe condition
②
Use this menu cell to enter the number of pipe diameters between the upstream configuration and the transducer
installation
5. Press <Right Arrow> key to select "=" sign.
6. Press <Up/Down Arrow> keys to select number of pipe diameters.
7. Press <ENTER> key to save selection.
8. To return to the top menu level, continue to press the <Left Arrow> key.
Table 3- 2
Pipe Configuration Option List Definitions
Fully Developed
1 Elbow
Dble Elbow+
Dble Elbow-
Valve
Expander
Reducer
Norm Entry
Header Inlet
Intrusions
Transducer Selection - Example for D1H High Precision Transducer
1. At the [Chan/Path Setup] menu, press <Down Arrow>key to select [Pick/install Xdcr].
2. Press <Right Arrow> key then <Down Arrow> key to [Transducer Model].
3. Press <Right Arrow> key.
20
upstream of your transducer mounting location. (Refer to the definitions below.)
Options
Fully developed flow, as would be expected for very long
straight pipe runs or installation downstream of a flow
condition.
Single 90 degree Elbow upstream of transducer installation.
Double out-of-plane Elbows upstream of transducer
installation.
Double in-plane Elbows upstream of transducer installation.
Not available at this time.
Pipe expansion upstream of transducer installation.
Pipe reduction upstream of transducer installation.
Not available at this time.
Header or pipe manifold upstream of transducer installation.
Not available at this time.
SITRANS FUS1020 IP65 NEMA 4 Quick Start
Operating Instructions (Compact), 4/2009, CQO:QSG006 Revision B
Definitions