5.1. HIGH PRESSURE SIDE OF THE MANIFOLD:
I1, I2 – INLET CONNECTIONS
The manifold contains left and right side threaded inlet connection that fits to the high pressure
hoses.
F1, F2 – FILTERS
The manifold contains two high pressure filters which are installed on the inlets of the manifold.
These filters can be easily maintained. Filters include antishock valves that prevent a backflow.
Antishock vlaves are not leaktight.
V1, V2 – INLET SHUT OFF VALVES
The manifold contains two inlet shut off valves that are used to open and close the main high
pressure on the left and right side of the manifold unit. By these two valves you can close gas
supply to the hospital pipeline.
V3, V4 – PURGE VALVES
The manifold contains two purge valves. Through the purge valves is ventilated high pressure
collecting pipelines on the empty side before removing high pressure hoses from the cylinders.
Also purge valves are used for cleaning of collecting pipeline before opening inlet shut off valve
and supply gas to the main pipeline. Outlet of purge valve has a threaded connection.
R1, R2 – REGULATORS
The manifold consists of two high pressure manifold regulators. The regulators reduce the
pressure from cylinder pressure to the outlet pressure setting on manifold outlet. Information
about setting of product you can find in testing report included to the product.