24. Go back one menu levels using [ESC].
Select "Low 0 %" and confirm with [↵].
25. Set the initial value of the volume measurement with respect
EN
to 0 % of the filling height (e.g. 0 % filling height corresponds
to 3 litres).
» The cursor moves to the next digit.
» Repeat for each digit.
» Initial value of volume measurement is set.
26. Select "High 100 %" and confirm with [↵].
27. Set the end value of the volume measurement with respect
to 100 % of the filling height (e.g. 100 % filling height
corresponds to 1,000 litres).
» The cursor moves to the next digit.
» Repeat for each digit.
» Initial value of volume measurement is set.
28 Go back two menu levels using [ESC].
Select "Mode" and confirm with [↵].
29. Go back one menu level using [ESC].
Select "Volume" and confirm with [↵].
» Mode is set to volume.
30. Perform a mounting correction.
→ See chapter 8.8 "Mounting correction (offset)".
» Volume measurement is configured.
32
WIKA operating instructions, model DMSU21A