DEBRIS BETWEEN BLADES
• Noise may occur if sand or debris are present between blades.
• The prosthetist should remove the Spectra Sock and foot cover from
the device, and use compressed air to remove sand or debris.
N.B.: The device should always be used with a foot cover and Spectra
Sock to minimize dust and dirt from entering between carbon blades.
LIABILITY
The manufacturer recommends using the device only under the specified
conditions and for the intended purposes. The device must be
maintained according to the instructions for use. The manufacturer is
not liable for damage caused by component combinations that were not
authorized by the manufacturer.
COMPLIANCE
ISO
- "P" - "m"kg *)
*)
Body mass limit not to be exceeded!
For speci c conditions and limitations of use see
manufacturer's written instructions on intended use!
safety checks. This device has been tested for moderate impact level use.
Refer to category selection chart for low and high impact level use. In the
standard mentioned, test levels (P) are assigned to certain maximal body
masses (m). In some cases, which are marked with*), no test level is
assigned to the device related to maximal body mass. In these cases, the
test loads have been adapted adequately on the basis of the specified
load level.
Category
1
2
3
4
5
6
7
8
9
WARRANTY
Össur provides a warranty of 36 months for the device and 6 months for
the foot cover.
ISO 10328 - label
Weight (Kg)
52
59
68
77
88
100
116
130
147
This device has been tested
according to ISO 10328 standard
to two million load cycles.
Depending on the user's activity
this corresponds to a duration of
use of two to three years. We
recommend carrying out yearly
Label text
ISO 10328–P3–52–kg
ISO 10328–P3–59–kg
ISO 10328–P3–68–kg
ISO 10328–P4–77–kg
ISO 10328–P4–88–kg
ISO 10328–P5–100–kg
ISO 10328–P6–116–kg
ISO 10328–P6–130–kg
ISO 10328–P6–147–kg