3. Tear open the pouch, remove the cartridge and discard the pouch
and desiccant pack.
4. From the Home screen on the TEG analyzer, select new qc.
5. Insert the cartridge into the slot, as indicated on the screen, with the
bar code on the left side.
6. On the Confirm Test screen, press continue.
7. After the cartridge pretest has completed and you have verified that
the assay is what you intended to run, press next. If the cartridge is
not the desired assay, press stop and start over.
8. On the Test Information screen, enter information for the test (if
desired), and then press next.
9. Pipette the prepared QC sample into the cartridge sample port, filling
up to or above the line marked on the cartridge. Press next.
The TEG analyzer starts the test, and the results are displayed as
they become available.
10. When the analyzer displays the "Remove cartridge" prompt, remove
the used cartridge from the slot and immediately dispose of it.
Expected Values
Expected values for test results when used with a Citrated Multichannel
Cartridge are within the reference ranges that were established based
on testing, following Clinical and Laboratory Standards Institute (CLSI)
publication C28-A3c, Defining, Establishing and Verifying Reference
Intervals. Each laboratory may establish its own level of performance to
monitor quality assurance.
Follow your institution's policy to establish mean target values and
ranges for the Abnormal QC material.