Installation and Maintenance

Installation may only be performed by qualified,
technical personnel in accordance with the service
instructions available from the manufacturer.
Note: Your supplier is informed and trained
about the scope and content of the installation
and maintenance work.
The light incl. support arm system must be
inspected annually by service personnel in
accordance with EN 62353 and, if necessary,
serviced. The inspections performed must be
documented and filed by the operator.
Attention: Before each use, check the system
incl. the handle for visible cracks or deformations
as well as for a firm and smooth-running
connection between the lamp head and the
support arm system!
Attention: Loose connections may cause the
lamp head to crash unexpectedly!
Do not operate the light if you have any doubts
about its mechanical stability, function or electrical
safety!