EMC Interference Immunity (general)
Please note that the light may only be
operated in an electromagnetic environment as
specified below (see table)!
Phenomenon
2
Discharge
3
Burst
4
Surge cable
Surge ground
6
Dip
Magnetic
7
field
8
HF fields I
9
HF fields II
If you have questions, contact your service technician!
1
Environment in home health care settings (including
professional health care facilities).
2
Static electricity discharge
3
Fast transient electrical disturbances
4
Surge voltage cable to cable
5
Surge voltage cable to ground
6
Voltage dips
7
Magnetic fields with energy technology rated frequencies
8
High-frequency electromagnetic fields
9
Conducted interference induced by high-frequency fields
Standard
section
IEC 61000-4-2
IEC 61000-4-5
IEC 61000-4-5
5
IEC 61000-4-5
IEC 61000-4-11
IEC 61000-4-8
IEC 61000-4-3
IEC 61000-4-6
Contact: ± 8 kV
Air: ± 2, 4, 8, 15 kV
± 2 kV @ 100 kHz
0% UT: 1 period
70% UT: 25/30 per.
0.08 – 2.7 GHz
0.15 – 80 MHz
41
1
Test level
± 1 kV
± 2 kV
30 A/m
10 V/m @
10 V
@
RMS