Methods; Chlorine Hr With Tablet - Aqualytic AL250 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Methods

GB
0.0.0
Zero
Test
CLHr
RESULT
20

Chlorine HR with Tablet

5 – 200 mg/l
Fill a clean vial (16 mm Ø) with 8 ml of the water sample and perform
zero calibration (see "Operation").
Remove the vial from the sample chamber.
Add one CHLORINE HR (KI) tablet straight from the foil to the water
sample and crush the tablet using a clean stirring rod.
Add one ACIDIFYING GP tablet straight from the foil to the same
water sample and crush the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl several times until the
tablets are dissolved.
Place the vial in the adapter making sure that the ı marks are aligned.
Place the cover on the adapter.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l Chlorine.
Tolerance: ± 5 mg/l