GE Baker Hughes PACE CM Instrucciones De Seguridad Y Guía Del Usuario página 13

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 41
1
11
1
Pressure source
4
Regulate to between 110% full-scale and MWP
6
Unit under test
9
Optional differential connection 
10 Vacuum generator
12 Exhaust to atmosphere
Figure 6: Pneumatic Connections with Negative Gauge Pressure Generator
Note: Refer to the PACE User Manual, Reference and Specification for details of other
system components.
*
High pressure gas exhaust - depending on pressure range.
** Optional vacuum system kit, allows the -ve port gas to be directly discharged to atmosphere, by-
passing the vacuum pump.
Optimum controller transient response and minimum time to set-point may be degraded if either the
pneumatic supply or vacuum system has restricted flow. Installing a reservoir volume, which has
larger capacity than the load volume, located in close proximity to the controller supply ports can
improve the controller response.
Optional negative gauge pressure generator kit, allows the -ve port to directly discharge to
atmosphere, by-passing the negative gauge pressure generator.
 To protect the control module (item 8), for ranges above 100 bar, from over-pressure a suitable
protection device (such as a relief valve or bursting disc) must be fitted to limit the applied supply
pressure to below the MWP.
 Optional differential connection kit.
© 2008 Baker Hughes, a GE company – All rights reserved.
3
4
2
7
10
12
2
Supply isolation valve
7
Optional reservoir †
11 Source pressure (regulated compressed air supply)
13 Check valve ‡
SUPPLY
+
-
8
7
14
3
5
8
a
14 Manual external vent valves
PACE CM Safety Instructions and User Guide-English | 9
OUTPUT
VENT REF
5
6
14
13
Filter
Diffuser *
Protection device 
Atmosphere
9
a