During use, only apply moderate force to the product. Otherwise,
parts of the product could break off and remain in the patient.
Avoid tissue contact with parts that are not insulated in the joint
area of the instrument. Otherwise, tissue in the area of the joint
can suffer thermal damage during activation.
Erbe Elektromedizin expressly warns against modifying the
product. Any modification exempts Erbe Elektromedizin from
any and all liability.
6
How to use
6.1
Only use sterile products
Clean, disinfect and sterilize these products before using them
for the first time and before every subsequent use.
6.2
Checking the product
1. Check the product for damage, including the insulation or
coating.
If damaged, do not use this product!
6.3
Assembling the product
CAUTION! Never press the instrument jaws together by hand to
prevent damaging the instrument.
1. To close the instrument jaws, hold the insert on the joint
sleeve (1) and pull on the rod (2).