Chattanooga Intelect Mobile 2 Guia De Inicio Rapido página 7

Ocultar thumbs Ver también para Intelect Mobile 2:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 62
• The Intelect® Mobile 2 may be susceptible to Electro-Static Discharge (ESD) at greater than ±6 kV when
first grasping the Ultrasound applicator� In the event of such a discharge, the Intelect® Mobile 2 may display
a permanent error� The Intelect® Mobile 2 will terminate all active outputs (stim, ultrasound,), automatically
place the unit in a safe state�
• To prevent Electro-Static Discharge (ESD) at greater than ±6 kV:
» Grasp and hold the Ultrasound prior to starting treatment� If the applicator must be put down prior to
completion of treatment, stop the current treatment first and then place the applicator in the holder�
» Maintain humidity in the use environment to at least 50% relative humidity�
» Floors should be wood, concrete, or ceramic tile� If floors are covered with synthetic material, DJO
recommends implementing additional controls to maintain relative humidity to at least 50%�
» Communicate these ESD-precautionary procedures to healthcare staff, contractors, visitors, and
patients�
WARNING
E N
7

Publicidad

Tabla de contenido
loading

Tabla de contenido